Description
The universal meter with TRUE RMS function and automatic range selection is an advanced measuring tool designed for precise diagnostics of electrical installations, electronic circuits, and industrial devices. Counting up to 60000 digits allows for accurate readings even of small changes in parameters, and the wide range of functions enables both basic and advanced measurements of electrical quantities.
The large, multifunctional LCD display, capable of simultaneously showing two measurement values, automatic screen backlighting, and illuminated measurement sockets significantly improve work comfort in challenging lighting conditions. A range of functions that facilitate measurements and wide measurement ranges, along with speed and industrial precision, will be appreciated by electricians, electronics specialists, and DIY enthusiasts alike. The sturdy housing with a non-slip protective PVC holster enhances user safety and protects the device from slipping out of hands and mechanical damage, while silicone-insulated measuring cables ensure high durability and resistance to elevated temperatures.
Basic Measurements:
- AC voltage: A.C. 0-750 V, TRUE RMS (40-1000 Hz)
- DC voltage: D.C. 0-1000 V
- AC current: A.C. 0-10 A, TRUE RMS (40-1000 Hz)
- DC current: D.C. 0-10 A
- Resistance: 0-60 MΩ
- Diode test: IF 1.8 mA, UR 3.2 V
- Continuity test: 0-50 Ω
Additional Measurements:
- Capacitance: 0-60 mF
- Frequency: 0-25 MHz
- Temperature: -20 to 1000°C
- Non-contact AC voltage detection: NCV
- Peak voltage measurement: PEAK VALUE
Measurement Accuracy:
Example of determining accuracy for AC voltage measurement. Calculations for measurement result: 235 V
- The accuracy of the YT-730896 meter for this range is: ±(0.8% + 10), resolution 0.01 V
- Error calculation formula: (measurement result x 0.8%) + 10 x resolution
- Calculations: (235 x 0.8%) + 10 x 0.01 = 1.88 + 0.1 = 1.98 V
The actual result considering the measurement error for the reading of 235 V = 235 ± 1.98 V
- Result considering the measurement error for the DC voltage reading: 12 V = 12 ± 0.011 V
- Result considering the measurement error for the AC current reading: 8 A = 8 ± 0.17 A
A table with resolution and accuracy for a given measurement can be found in the user manual.
Functions:
- TRUE RMS - the meter displays the true RMS value of AC voltage/current; unlike averaging meters, devices with TRUE RMS show the correct value not only for sinusoidal waveforms but also for distorted sinusoidal waveforms, e.g., higher harmonics or square and triangular waveforms; a must-have function wherever switching power supplies, motor drive inverters, HVAC systems, etc., are present in the circuit.
- NCV - non-contact AC voltage detection using induction; the meter detects the presence of voltage in an insulated wire using the NCV sensor on the front of the device.
- AUTO - automatic range adjustment (only one measurement needed), activated by default; in the case of measuring unknown values using a meter with manual range adjustment via a dial, measurements should be taken sequentially from the largest range down to the range providing the most accurate result - otherwise, there is a risk of damaging the device.
- RANGE - manual range adjustment, useful when checking multiple identical components with known values, e.g., a series of resistors; manually setting the range will significantly shorten the time for such measurements.
- REL - using this function during measurement will zero the display and take the visible value before displaying as the reference level; the next measurement will show the difference between the measured value and the stored reference value.
- FUNC - changing the measurement mode, e.g., from AC voltage measurement to DC.
- MAX/MIN - holding the maximum or minimum value from the current measurement on the display.
- PEAK - peak value for AC current and voltage measurements.
- HOLD - holding the currently measured value on the display.
Additional Information:
- LCD display, maximum displayed result: 60000
- Screen backlighting with a light sensor, automatically illuminates the screen depending on ambient brightness.
- Measurement sockets illuminate when a specific measurement is selected (clear hint on which measurement sockets to insert the probes for that measurement).
- Audible and visual signal in the continuity test for resistance below 50 Ω.
- Ability to turn on diode and continuity tests separately.
- Measurement probes 13 cm long.
- Length of the cable with probes: 130 cm.
- Cable diameter with insulation: 3.5 mm.
- Point probe for temperature measurement.
- Design with overload protection.
- ABS plastic housing.
- Automatic shutdown of the meter after 15 minutes of inactivity.
- Low battery indicator.
- Soft case with zipper.
- Power supply: 3 x 1.5 V batteries (AAA).
| ◀ Article | Parameters | In stock | Price: | ▶ | |
|---|---|---|---|---|---|
| YT-730896 | Alternating voltage : 0-750 V a.c., Direct voltage: 0-1000 V d.c., Alternating current: 0-10 A, Direct current: 0-10 A, Resistance: 0-60 Mω, Capacity: 0-60 mF, Frequency: 0-25 Mh, Z, Temperature: -20 ~ +1000 ˚C, Continuity test: 0-50 Mω, Conduction test: yes, Diode test: yes, Minimum measurement value: yes, Maximum measurement value: yes, Relative value: yes, Peak value: yes, True Rms: yes, Data Hold: yes, Non-contact Ac voltage detection Ncv: yes, Dimensions: 187x95x55 mm, Material: Abs, Display type: Lcd, Led backlight: yes, Display range: 60000, Auto measurement range: yes, Sampling frequency: 3 times per second, Bar graph: 0-100, Bell ring: yes, Auto power off: 15 min, Front lighting: no, Low battery indicator: yes, Power supply: battery, Battery type: 3x1.5V, Aaa, Working temperature range: 0~40 ˚C, Operating relative humidity Rh:, 23 × 14 × 7 cm, 0.538 kg | Available | 41.12 EUR | Request |